MTDC-64 is a 64-channel, 100 ps time-stamping TDC with PCI-Express interface. Compact and portable lab setups may be built with this device.
Time-Of-Flight spectroscopy in nuclear physics is available with MTDC-64 device. It is designed for production testing of particle detectors: straw tubes, gas-filled drift chambers, scintillator counters, neutron counters. Tasks which involve medium channel count in compact size no more needs huge racks at high cost. Front-end electronics, data readout and storage are accomplished inside one PC now.
Features
- 64 channels per PCI slot
- up to 6 boards, 384 channels in one PC
- 100 ps time bin
- Time resolution: 64 ps RMS, 34 ps RMS with correction
- 6.0 ns minimum pulse width
- 8 MHz maximum hit rate per channel
- 2048 measurements per trigger
- External or Self-Trigger
- External or Timer start of run
Interface
- PCI-Express 1x interface
- 64 LVDS/dECL hit inputs
- NIM inputs: Start Run, Trigger
- 32 front panel inputs, 32 on side
- synchronization and extension connector on rear
User Manuals
Understanding Trigger Matching
Test Reports
Interface
- NIM input CL2 (upper) - trigger signal, 6.0 ns minimum pulse width (30 ns in firmware prior to r5814)
- NIM input CL1 (lower) - spill start signal, 55 ns minimum pulse width
- 68-pin connector P1: input channels 1-32
- Internal two 34-pin connectors: P2(channels 33-48), P3 (49-64)
- Green LED D2: data taking in progress
- Red LED D1: Error or Not-Configured status
Internal TTC bus on P5
Software
Control, histogramming and spectra analysis provided with VMEDAQ software.
Registers
Detailed description of MTDC-64 Registers